GasEye Extractive

The GasEye Extractive is a versatile gas analyzing tool for industrial process applications. A process gas sample is continuously fed into the analyzer where it is analyzed in real-time utilizing laser absorption spectroscopy.

It can be configured to operate in the near-infrared (NIR), mid-infrared (MIR), and infrared (IR) wavelength range thereby allowing to analyze the majority of gases of interest in the industrial process monitoring.

GasEye Extractive analyzer is assembled in a 19” RACK IP20 or wall-mounted IP66 enclosure.

Applications
  • Safety monitoring
  • Process Control
  • Custody transfer.
Industries
  • Chemical industry
  • Petrochemical industry
  • Food industry.
Features
  • Sample temperature: 0 – 200°C
  • Real time sensing: response time below 1s
  • High selectivity: automatic compensation for interference effect from other constituents in the gas sample
  • High sensitivity: detection limits down to ppb range
  • Maintenance free: equipped with a self-calibrating feature, no field calibration necessary
  • ATEX version available
Most common application
Model: Minimum range: Typical Application:
GasEye Extractive H2O 0 – 5 ppmv Moisture in NG, LNG, H2
GasEye Extractive H2S 0 – 10 ppmv H2S in NG, LNG, H2
GasEye Extractive H2 0 – 10 vol% Hydrogen applications
GasEye Extractive CO/CO2/CH4 CO: 0 – 1 ppmv
CO2: 0 – 2 ppmv
CH4: 0 – 10 ppmv
Hydrogen purity PSA/PSA+
GasEye Extractive HCOOH/HCHO/NH3 HCOOH: 0 – 1 ppmv
HCHO: 0 – 1 ppmv
NH3: 0 – 1 ppmv
Hydrogen purity
GasEye Extractive H2/O2 H2: 0 – 10 vol%
O2: 0 – 100 ppmv
Hydrogen applications
Operating principle

GasEye Extractive is a laser spectrometer that uses single line molecular absorption spectroscopy. A semiconductor laser emits a beam of infrared light which passes through the process and impinges the detector on the receiver side. When the target gas is present the light intensity changes and this can be used to determine the concentration of the measured specimen. The wavelength of the radiation emitted by the laser is chosen to match the specific absorption line wavelength of the gas. The laser continuously scans this single absorption line with a very high spectral resolution. The measurement is free of cross-interferences since the laser light is absorbed very selectively by only one specific line in the scanned spectral range.

19″ RACK VERSION
 

GasEye Extractive 19″ RACK analyzer is assembled in a standard 19-inch rack 4U version.

It exists in several various configurations from single gas to multi gas applications.

Specification:
Dimensions: width x height: 482 mm x 177 mm
length: 650 mm
Weight: 35 kg (for the model with multipass cell, vacuum pump, and safety valves)
Gas connection: 6 mm tubing
Degree of protection: IP20 to EN 60529
Materials:
Housing: aluminium
Internal process cell: stainless steel 316
Cell window: sapphire
 Climatic conditions
Ambient temperature: 5ºC to 40ºC
Ambient pressure: 800 – 1200 hPa
Ambient humidity: RH < 90%, non-condensing
Measurement conditions
Sample gas pressure: 0.05 – 5 barG
Sample gas temperature: ambient
Sample gas flow: 0.5 – 20 Ndm3/min
Safety:
Low Voltage Directive (LVD) 2014/35/EU PN-EN 60825-1:2014-11 – Safety of laser products – Part 1: Equipment classification and requirements
PN-EN 61010-1:2011 – Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements
EMC Directive 2014/30/EU EN 61326-1:2013-6 – Electrical equipment for measurement, control and laboratory use
RoHS Directive 2011/65/EU

 

WALL MOUNTED VERSION

The GasEye Extractive Wall Mounted type analyzer exists in general purpose and hazardous area configuration (version Ex1). It exists in several various configurations from single-gas to multi-gas applications. The system is IP66 rated with ambient temperature ranges from -30ºC to +60ºC and can be mounted outside without a need for a container.

GasEye Extractive Ex1 provides Ex p (purge and pressurization) type of explosion protection. Ex p pressurized equipment is the technique of guarding against the ingress of the external atmosphere into an enclosure by maintaining a protective gas therein at a pressure above that of the external atmosphere. The overpressure is maintained either with or without a continuous flow of the protective gas.

 

Specification:
Dimensions: width x height: 460 mm x 230 mm
depth: 650 mm
Weight: 60 kg
Gas connection: connector 6 mm Swagelok
Materials:
Housing: SS 1.4301 (AISI 304)
Internal process cell: SS 1.4404 (AISI 316L)
Cell window: sapphire
Climatic conditions:
Ambient temperature: -30ºC to +60ºC
Ambient pressure: 800 – 1200 hPa
Ambient humidity: RH < 90%, non-condensing
Measurement conditions:
Precision: +/- 1% of the measured value or LOD and response time 30 sec
Accuracy: +/- 2% of the measured value or LOD and response time 30 sec
Conditioning of the sample: dry, free from oils and particles
Calibration: factory calibration using certified gas
Zero drift and span drift: irrelevant
Sample temperature: 0°C to 200°C (depends on application)
Safety:
Low Voltage Directive (LVD) 2014/35/EU PN-EN 61010 – 1:2011- Safety requirements for electrical equipment for measurement, control and laboratory use
PN-EN 60825 – 1:2014-11- Safety of laser products
EMC Directive 2014/30/EU EN 61326-1:2013-06 – Electrical equipment for measurement, control and laboratory use
RoHS Directive 2011/65/EU
EN IEC 60079-0:2018 – Explosive atmospheres – Part 0: Equipment – General requirements
EN 60079-2:2014 – Explosive atmospheres – Part 2: Equipment protection by pressurized enclosure “p”
EN 60079-26:2015 – Explosive atmospheres – Part 26: Equipment with Equipment Protection Level (EPL) Ga
EN 60079-28:2015 – Explosive atmospheres – Part 28: Protection of equipment and transmission systems using optical radiation
ATEX Directive 2014/34/EU  
Explosion protection (standard version):

  • ATEX II 3G [Ex op is IIC T6 Gc]
  • ATEX II 3D [Ex op is IIIC T85°C Dc]
Explosion protection – ATEX Zone 1/21 (optional version):

GasEye Extractive Ex1:

  • II 1/2G Ex op is pxb IIC T* Ga/Gb
  • II 1/2D Ex op is pxb IIIC T** Da/Db

GasEye Extractive Ex1 ET:

  • II 1/2G Ex db eb h ia ib op is pxb q IIC T* Ga/Gb
  • II 1/2D Ex h ia ib op is tb pxb q IIIC T** Da/Db

*/** temperature class depends on the medium temperature according to:

ID Medium temperature: Ambient conditions: Temperature class: Maximum surface temperature:
1 < 50°C -30°C to 50°C T6 85°C
2 < 100°C -30°C to 60°C T5 100°C
3 < 135°C -30°C to 60°C T4 135°C
4 < 200°C -30°C to 60°C T3 200°C

Certificate No. KDB 20ATEX0056X EN
Certificate No. KDB 20ATEX0056X PL

Brand

Stack, Process & CEMS

Categories , Tags , , , , , , , , , Brand:

Request information or pricing

If you need a quote, specifications or technical info, or anything else – please submit the form below and one of our team of experts will happily get back to you within 24 business hours. 

  • This field is for validation purposes and should be left unchanged.

Related Products

.woocommerce .woocommerce-tabs h2 { display: none; }